The local monitoring and alarm system, MGSAP L6, provides the clinical emergency alarms (clinical emergency alarms) in compliance with EN 7396-1, HTM 02-01 and EN 737-3 and consists of an electronic board that continuously monitors the pressure of medical gases by means of pressure transmitters and/or pressure switches.
Features:
- Microprocessor based design providing powerful functionality
- Built in LCD screen (2×16 characters) for displaying all measurements
- Power supply of 230V 50/60 Hz or 24V DC
- Audible signal for High/Low alarm condition
- Up to 9 Analog Inputs (6 provided as standard and 3 as option), 4-20 mA range, 10 Bit resolution
- Up to 12 independent Digital Inputs (pressure switches and dry contacts)
- Alarm/event report and log file (up to 100 records)
- Built in RJ45 port for Ethernet connection (1 port, 10/100 Base-T) for web interface (as standard)
- MODBUS TCP/IP supported for connection to BMS systems (as standard)